7 - settings submenu – Hudson Robotics Micro10x User Manual

Page 15

Advertising
background image

15

Figure 2.6.1

Time – This is the interval in seconds; a time of 0 deactivates this feature

Volume – This is the total volume that will be moved through the manifold. Enter
a value between 120 and 9999uL.

Speed – This adjusts the percentage of the pumps maximum speed. More viscous
liquids will require a slower pump speed to avoid cavitation. Do not exceed more
than 30% speed on any value over 1000ul.

2.7 - Settings Submenu


There are critical settings that can be adjusted by an administrator (Figure. 2.7.1).
When you access this submenu you will be ask a question please click yes to
continue to the Settings menu (Figure 2.7.2)