Section 3 - maintenance of the mirco10x fluid path, 1 - removal of the pump head – Hudson Robotics Micro10x User Manual

Page 22

Advertising
background image

22

Section 3 - Maintenance of the Mirco10X Fluid Path


Note: Please turn power off and disconnect all cables from the rear of the Micro10x
before performing any maintenance on the unit.



3.1 - Removal of the Pump Head


Remove the 4 mounting screws from the face of the pump (Figure 3.1.1). Support
the pump with your hand to prevent it from falling from the mounting plate.

Figure 3.1 1

Gently pull the pump diagonally away from the mounting plate. Care must be
taken not to pull the piston more than 1” out of the cylinder; damage to the piston
seals may occur (Figure 3.1.2). Once the pump head is far enough away from the
pump module you will slowly lift up on the entire pump assembly. You will need
to remove the piston drive shaft from the pivot bearing inside the pump module
(Figure 3.1.3).

4 mounting
screws for
Pump head

Advertising