Section 1 - Setup

1.1 - Tubing

The MC10X is shipped with two tube assemblies (Figure 1.1.1). The tube
assembly with the two Flangeless fittings should be installed between the right
port on the pump head and the MC10X manifold. To create a proper seal tighten
the fittings into the pump head and manifold until they stop finger tight. Once the
fitting is finger tight turn the fitting an additional 1/8 turn to ensure a proper seal.
(Figure 1.1.2)

Install the longer tubing assembly in the same manner; with the single Flangeless
fitting between the left port on the pump head and your reagent reservoir. (Figure
1.1.3)

Figure 1.1.1

Tubing
Pump to
Reservoir

Tubing Pump to Manifold