plate nest without touching it. To adjust, place a sheet of paper or business card
on the plate nest. Jog the manifold down until it touches the paper but not pinch it
to the nest. (Figure 2.7.6)

Figure 2.7.6

Adjust Plate Origin – This value should be evaluated and adjusted if manifold
types or plate nests are changed. This is the reference point for all well positions
in a SBS standard plate. This is taught by aligning tip 1 on the manifold with the
intersection of wells A1, A2, B1 and B2 in a 384 well plate. Tips 2-12 should
align with every other well intersection in that row. (Figure 2.7.7)