2.6 – Change Settings

This Section allows you to modify all the performance settings of the RapidWash
(Figure 19)

Figure 19

Settings Tab:

Aspirate Shift-

This is the distance that the RapidWash will use above and in the

plate wells to begin the aspirate step. +mm is above the top of the Plate and –mm
is aspirate begins in the plate.


Dispense Shift-

This is the distance that the RapidWash will use above and in the

plate wells to begin the dispense step. +mm is above the top of the Plate and –mm
is aspirate begins in the plate.

Travel Shift-

This is the distance that the RapidWash will travel above the Plate

when the manifold moves between Aspirate, Dispense, and Prime Positions. +mm
is above the plate


Teaching Plate-

This is the plate height of the microtiter plate that is used on the

calibration procedures (a standard 96 shallow well SBS Plate is usually between
13.5mm-14.0mm).

Dwell Time-

This is the delay time the aspirate head stays in the plate after it stops

at the defined depth location. This setting is set in seconds default is 0.5 seconds.


Prime Volume-

This is the amount of fluid in uL the RapidWash primes when the

unit switches it liquid path during a wash program.