Hudson Robotics RapidWash User Manual

Page 19

Advertising
background image

19

Figure 21


Aspirate Origin-

This is the saved position of the aspirate manifold when the

RapidWash begins to aspirate. This position is calibrated with a 96 well teaching
plate. The position should be taught where each set of tips are centered in each
well of the plate (Figure 22).

Figure 22