3 serial command calibration – Rice Lake 920i Installation Manual V2.0 User Manual

Page 51

Advertising
background image

Calibration

47

To calibrate the indicator using the front panel, do the
following:

1. Place the indicator in setup mode (display

reads

Scale Configuration

) and remove all

weight from the scale platform. If your test
weights require hooks or chains, place the
h o o k s o r c h a i n s o n t h e s c a l e f o r z e r o
calibration.

2. With the

SCALES

menu highlighted, press the

d o w n

key, t h e n s e l e c t t h e s c a l e t o b e

c a l i b r a t e d . P r e s s

d o w n

a g a i n (

G R A D S

parameter highlighted), then press

left

to

h i g h l i g h t t h e

C A L I B R

s u b m e n u ( s e e

Figure 4-1). Press

d o w n

to go to zero

calibration (

WZERO

). See Figure 4-2 on page

46.

3. Ensure scale is empty, then press

down

again

to show the current WZERO value, then press
the

Calibrate

softkey to calibrate zero. When

complete, the new A/D count for the zero
calibration is displayed. Press

enter

to save

the zero calibration value and go to the next
prompt (

WVAL

).

4. With

WVAL

displayed, place test weights on

the scale and press

down

to show the test

weight value. Use the numeric keypad to enter
the actual test weight, then press

enter

to save

the value and go to span calibration (

WSPAN

).

5. Press

down

again to show the current

WSPAN value, then press the

Calibrate

softkey to calibrate span. When complete, the
new A/D count for the span calibration is
displayed. Press

enter

again to save the span

calibration value and go to the next prompt
(

WLIN

).

6. Five-point linearization (using the WLIN

parameter) provides increased scale accuracy
by calibrating the indicator at up to Þve
additional points between the zero and span
calibrations.
Linearization is optional: if you choose not to
perform linearization, skip the WLIN
parameter; if linearization values have
previously been entered, these values are reset
to zero during calibration. To perform
linearization, follow the procedure below:
With

WLIN

displayed, Press

down

to go to the

Þrst linearization point (

POINT 1

). Press

down

again to show the weight value prompt

(

WGT 1

), then

down

once more to show the

weight value. Place test weights on the scale,
then use the numeric keypad to enter the
actual test weight value. Press

enter

to save

the value and move to the calibration (

CAL 1

)

prompt. Press

down

to show the current

calibration value, then press the

Calibrate

softkey to calibrate the linearization point.

When complete, the A/D count for the linear
calibration is displayed. Press

enter

again to

save the calibration value and go to the next
prompt (

POINT 2

).

Repeat for up to Þve linearization points. To
exit the linearization parameters, press the

up

key to return to WLIN.

7. The optional rezero function is used to

remove a calibration offset when hooks or
chains are used to hang the test weights.

NOTE:

The rezero function cannot be used

with five-point linear calibration.

¥

If no other apparatus was used to hang the
test weights during calibration, remove
the test weights and press

up

to return to

the CALIBR submenu.

¥

If hooks or chains were used during
calibration, remove these and the test
weights from the scale. With all weight
removed, go to the REZERO parameter,
then press

down

to show the current zero

value. Press the

Calibrate

softkey to

adjust the zero and span calibration
values. Press

enter

or

up

to return to the

CALIBR submenu.

8. Press

up

to return to the SCALES menu, or

press the

Save and Exit

softkey to exit setup

mode.

4.3

Serial Command Calibration

To calibrate the indicator using serial commands, the
indicator serial port must be connected to a terminal
or personal computer. See Section 2.3.3 on page 7 for
serial port pin assignments; see Section 9.0 on
page 78 for more information about using serial
commands.

Once the indicator is connected to the sending device,
do the following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from

the scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.

2. Send the SC.WZERO#n serial command

(where n is the number of the scale) to
calibrate zero.

3. Place test weights on the scale and use the

SC.WVAL command to enter the test weight
value in the following format:

SC.WVAL#n=vvvvv<CR>

4. Send the SC.WSPAN#n serial command to

calibrate span.

5. Up to Þve linearization points can be calibrated

between the zero and span calibration values.

Advertising