• For load cell cables, cut the shield wire just

past the grounding clamp. Shield wire

function is provided by contact between the

cable shield and the grounding clamp.

• Route stripped cables through cord grips and

clamps. Ensure shields contact grounding

clamps as shown in Figure 2-1. Tighten

grounding clamp nuts.

• Finish install ation using cable ti es to se cure

cables inside of indicator enclosure.

2.3.2

Load Cells

To attach cable from a load cell or junction box to an

installed A/D card, route the cable through the cord

grip and ground the shield wire as described in

Section 2.3.1 on page 8.
Next, remove connector J1 from the A/D card. The

connector plugs into a header on the A/D card (see

Figure 2-2). Wire the load cell cable from the load cell

or junction box to connector J1 as shown in Table 2-1.

Figure 2-2. Single-Channel A/D Card

If using 6-wire load cell cable (with sense wires),

remove jumpers JP1 and JP2 before reinstalling

connector J1. For 4-wire installation, leave jumpers

JP1 and JP2 on. For 6-wire load cell connections on

dual-channel A/D cards, remove jumpers JP3 and JP4

for connections to J2.

When connections are complete, reinstall load cell

connector on the A/D card and use two cable ties to

secure the load cell cable to the inside of the

enclosure.

S

IG+

S

IG–

S

EN+

S

EN–

EXC+

EXC–

J1

JP2

JP1

A/D Card

Connector Pin

Function

1

+SIG

2

–SIG

3

+SENSE

4

–SENSE

5

+EXC

6

–EXC

For 6-wire load cell connections to connector J1, remove
jumpers JP1 and JP2.

For 6-wire load cell connections to connector J2 (dual A/D
cards), remove jumpers JP3 and JP4.

Table 2-1. A/D Card Pin Assignments