Rice Lake 920i Installation Manual V3.13 User Manual

Page 59

Advertising
background image

Calibration

53

Figure 4-2. WZERO Calibration Display

To calibrate the indicator using the front panel, do the

following:

1. Place the indicator in setup mode (dis play

reads

Scale Configuration

) and remove all

weight from the scale platform. If your test

weights require hooks or chains, place the

hooks or chains on the scale for zero

calibration.

2. With the

SCALES

menu highlighted, press the

d o w n

key, then se lect the scale to be

calibrated. Press

d o w n

again (

G R A D S

parameter highlighted), then press

left

to

highlight the

C A L I B R

submenu (se e

Figure 4-1). Press

d o w n

to go to zero

calibration (

WZERO

). See Figure 4-2 on

page 53.

3. Ensure scale is empty, then press

down

again

to show the current WZERO value, then press

the

Calibrate

softkey to calibrate zero. When

complete, the new A/D count for the zero

calibration is displayed. Press

enter

to save the

zero calibration value and go to the next

prompt (

WVAL

).

4. With

WVAL

displayed, press

down

to show the

stored calibration weight value. Use the

numeric keypad to enter the actual value of

the calibration test weights, then press

enter

to

save the value and go to span calibration

(

WSPAN

).

5. Place test weights on scale. Press

down

again

to show the current WSPAN value, then press

the

Calibrate

softkey to calibrate span. When

complete, the new A/D count for the span

calibration is displayed. Press

enter

again to

save the span calibration value and go to the

next prompt (

WLIN

).

6. Five-point lin earization (us ing the WLIN

parameter) provides increased scale accuracy

by calibrating the indicator at up to five

additional points between the zero and span

calibrations.
Linearization is optional: if you choose not to

perform linearization,

skip the WLIN

parameter; if l

inearization val ues have

previously been entered, these values are reset

to zero during calibration. WZ

ERO and

WSPAN must be calibrated before adding

linearization points; WLIN val ues must be

less than the WSP AN va lue and must not

duplicate WZERO or WSPAN.
To perform linearization, follow the

procedure below:
With

WLIN

displayed, Press

down

to go to the

first linearization point (

POINT 1

). Press

down

again to show the weight value prompt (

WGT

1

), then

down

once more to show the weight

value. Place test weights on the scale, then use

the numeric ke ypad to enter the ac tual test

weight value. Press

enter

to save the value and

move to the calibration (

CAL 1

) prompt. Press

down

to show the current ca libration va lue,

then press the

Calibrate

softkey to calibrate the

linearization point. When complete, the A/D

count for the li near calibration is displayed.

Press

enter

again to save the calibration value

and go to the next prompt (

POINT 2

).

Repeat for up to five linearization points. To

exit the linearization parameters, press the

up

key to return to WLIN.

7. The optional rezero

function is used to

remove a calibration offset when hooks or

chains are used to hang the test weights.

NOTE: The rezero function cannot be used with
five-point linear calibration.

• If no other apparatus was used to hang the

test weights during calibration, remove

the test weights and press

up

to return to

the CALIBR submenu.

• If h ooks or chain s were used du ring

calibration, remove these and the test

weights from the scale. With all weight

removed, go to the REZERO parameter,

then press

down

to show the current zero

value. Press the

Calibrate

softkey to adjust

the zero and span calibration values. Press

enter

or

up

to return to the CALIBR

submenu.

8. Press

up

to return to the S CALES menu, or

press the

Save and Exit

softkey to exit setup

mode.

07/14/2006

01:57PM

Calibrate

WZERO

WVAL

REZERO

1291

Deadload Calibration

+/-

Millivolts

Zero Count =>

Last Zero

Temp Zero

CALIBR

SCALE1