Configuration

25

DIGFLT1
DIGFLT2
DIGFLT3

4
8
16
32
64
128
256
1
2

Selects the digital filtering rate used to reduce the effects of mechanical vibration from the
immediate area of the scale.

Choices indicate the number of A/D conversions per update that are averaged to obtain the
displayed reading. A higher number gives a more accurate display by minimizing the effect
of a few noisy readings, but slows down the settling rate of the indicator.

NOTE: When configuring non-A/D scales, set the DIGFLTx parameters to 1 to disable
filtering.

See Section 10.9 on page 116 for more information about digital filtering.

DFSENS

2OUT
4OUT
8OUT
16OUT
32OUT
64OUT
128OUT

Digital filter cutout sensitivity. Specifies the number of consecutive readings that must fall
outside the filter threshold (DFTHRH parameter) before digital filtering is suspended.

See Section 10.9 on page 116 for more information about digital filtering.

DFTHRH

NONE
2D
5D
10D
20D
50D
100D
200D
250D

Digital filter cutout threshold. Specifies the filter threshold, in display divisions. When a
specified number of consecutive scale readings (DFSENS parameter) fall outside of this
threshold, digital filtering is suspended. If NONE is selected, the filter is always enabled.

See Section 10.9 on page 116 for more information about digital filtering.

RATTRAP

OFF
ON

Enables RATTLETRAP

®

digital filtering. RATTLETRAP is most effective at filtering repeating

vibrations caused by mechanical noise from nearby machines but may increase settling
times over standard digital filter selections.

SMPRAT

30HZ
60HZ
120HZ
240HZ
480HZ
960HZ
7.5HZ
15HZ

Sample rate. Selects measurement rate, in samples per second, of the analog-to-digital
converter. Lower sample rate values provide greater signal noise immunity.

NOTE: The maximum total sample rate for all configured A/D channels—the sum of the
sample rates for all scales—is 1200 Hz. For example, up to ten scales can be configured
with 120 Hz sample rates, or up to twenty scales with 60 Hz sample rates.

PWRUPMD

GO
DELAY

Power up mode. In GO mode, the indicator goes into operation immediately after a brief
power up display test.

In DELAY mode, the indicator performs a power up display test, then enters a 30-second
warm up period. If no motion is detected during the warm up period, the indicator becomes
operational when the warm up period ends; if motion is detected, the delay timer is reset
and the warm up period repeated.

TAREFN

BOTH
NOTARE
PBTARE
KEYED

Enables or disables push-button and keyed tares. Possible values are:

BOTH: Both push-button and keyed tares are enabled
NOTARE: No tare allowed (gross mode only)
PBTARE: Push-button tares enabled
KEYED: Keyed tare enabled

ACCUM

OFF
ON

Accumulator. Specifies whether the scale accumulator is enabled. If enabled, accumulation
occurs whenever a print operation is performed.

VISIBL

ON
OFF

Scale visibility. Specifies whether scale data is displayed.

SCALES Menu

Parameter

Choices

Description

Table 3-2. SCALES Menu Parameters (Continued)