Calibration

55

4. Remove all weights from the scale. If chains

or hooks are used, place them on the scale.

Press the

Calibrate Zero

button to perform the

zero calibration (see Figure 4-6). A message

box appears when the process is complete.

Figure 4-6.

iRev Zero Calibration Display

5. Apply the test we ights to the scale. Press the

Calibrate Span

button to perform the span

calibration (see Figure 4-7). A message box

appears when the process is complete.

Figure 4-7.

iRev Span Calibration Display

6. If the option for chains or hooks was selected

in step 3, the Rezero display is shown (see

Figure 4-8). Remove all weights from the

scale, including chains or hooks. Press the

Re-Zero

button to calibrate the zero offset.

Figure 4-8.

iRev Rezero Display

7. If performing a multi-point li near calibration,

up to five other calibration weight values can

be entered on the display shown in Figure 4-9.

The weights must be in ascending order and

must not include zero or the span weight.

Enter the weight values and click the

Go

button to calibrate each point.

Figure 4-9.

iRev Linear Calibration Display

8. Review the new calibration values then click

Finish

to close the Cal ibration Wizard. To

restore the current calibration values, click

Cancel

.

Figure 4-10.

iRev Calibration Values Display