Maintenance 23

5.0

Maintenance

5.1 Troubleshooting

Symptom

Cause

Possible Solution

There is no indication on the display

The AC adapter is not connected

Connect the AC adapter

The display is unstable, [M] remains
flashing without charging

The balance is subject to air currents or vibration

See Section 1.1 on page 1.

The balance is situated on an unstable surface

An object is contacting the sample being measured,
the measuring pan, or the tare

Weight indication contains an error

An error was made in the tare subtraction procedure

Review the tare subtraction

The adjusters remain lifted, resulting in an incorrect
level

Check the level

The indication values are inconsistent after long hours
of use, or after being moved to a new location

Execute span adjustment on the
balance

The limit function does not work

The limit function is not selected

Check the operation of th elimit
function

The limit value has been entered incorrectly

[



] appears ([

] and a value flash

at [LO]).

Likely to produce errors in the counting mode
because the sample weight is insufficient

Execute the memory update
method

[



] appears before the

capacity is reached

Gross weight exceeded the capacity of the balance
(weight range= container + weight of sample)

Check the total weight, execute
tare subtraction again

A section of the mechanism is damaged

Contact our technical service
division

[u-Err] is displayed

A foreign object is caught between the measuring
pan (pan base) and the balance

Remove the measurement pan
and examine the surface
beneathe it

A section of the mechanism is damaged

[b-Err] is displayed
[d-Err] is displayed

The balance is exposed to static electricity or noise

Contact or technical service
division

The electrical system of the balance is malfunctioning

During span ad justment (TP):
[0-Err] is displayed
[1-Err] is displayed
[



] is displayed

A weight heavier than the capacity was used

Check that the span adjustment
procedure was performed
correctly

The reference weight is less than 50% of the capacity

Calibration produced an error of 1.0% or more

During battery installation the

indication disappears and [

]

flashes

The automatic power-off function was activated

Switch on the power again. If
necessary, deactivate automatic
power-off.

The battery capacity is low

Recharge the battery or operate
the balance with the AC adapter