28 TP Series Balance

5.3 NTEP Sealing and Operation

The balance is sealed by threading a wire through two tabs on the bottom front of the scale (see below-right image). One tab is

part of the top half of the enclosure and the other is part of the bottom half. There is also a flat plastic plug that covers a hole on

the bottom front of the balance which is sealed with a tamper-evident seal. If the plug is removed, a small switch can be

accessed to toggle between NTEP and non-NTEP modes.

When in NTEP mode, an outline of the least significant digit appears on the display as shown in the image below with digit 3.

NTEP mode does not allow access to the calibration mode.

Depending on capacity, there are two ways to seal the TP Series.

To seal the
balance, the lead
seal would be
placed here.

The tamper-proof
seal and plug deny
access to the NTEP
Mode toggle switch.

Tamper-proof seal
and plug removed for
access to the NTEP
Mode toggle switch.