4.0 Calibration

Enter the setup mode by pressing the internal CAL switch on the circuit board. Assume

for the following procedure that the desired capacity resolution is 5000lb by 1-lb. If

the selection for F5.0 above is ‘2’, lb/kg switchable, then enter all calibrations fields

below as pounds. The kg equivalent will be automatically computed during normal

scale operation.

DISPLAY

ACTION / DESCRIPTION

FSE F1

Advance to F step 5 by pressing ZERO.

FSE F5

Enter CAL mode by pressing PRINT.

F5.0 2

Use the TARE key to select 2 for calibration in pounds and press

PRINT to advance to the next step.

F5.1 1

Use the TARE key to display 1 for the increment size and press

PRINT to advance to the next step.

F5.2 0

Use the TARE key to display 0 for the decimal location on the display

and press PRINT to advance to next step.

F5.3 E

Use the TARE key to display ‘E’ to enter calibration and press PRINT

to advance to the next step.

F5.4 0

Use the TARE key to display 0 for no linearity correction and press

PRINT to advance to the next step.

E.SCALE

Remove all weight from the scale and press PRINT.

0

Enter the full scale capacity. Use the TARE to select the first digit

of the capacity then press GROSS NET to move it to the left. This

sequence would result in ‘50’.

5000 Repeat the process in the step above for the second digit until all digits

are entered then press PRINT.

07 SEC

The display counts down ensuring a stable reading.

Add Ld

Apply a test weight to the scale or full capacity if available and press

PRINT.