1 calibration error codes, 2 error displays – Rice Lake DWM-IV Digital Weighmeter Model 5 User Manual

Page 19

Advertising
background image

0.1 Calibration Error Codes

CE1 Insufficient test weight. Value must be greater than or equal to 5% of Full

Scale Capacity.

CE2 Full Scale Capacity will exceed internal A/D counts. Remedy by lowering

Full Scale Capacity weight (CAL step E. Scale.) Deadload may be

too high initially. Add a resistor across +SEN and –SIG on TB3 to

lower the counts.

CE3

Insufficient A/D counts for the resolution selected. Remedy by

increasing the increment size (Step F6.1.)

CE4

Motion detected during delay period. Ensure that the scale base is

not bumped or unstable during calibration.

CE5

The scale has gone into overcapacity, undercapacity, or negative

during delay period. Check load cell connections and polarity of

leads.

0.2 Error Displays

OLoAd

Applied weight has exceeded 102% of full scale capacity.

? ? ? ? ? ? Top dashes indicate that the maximum A/D counts have been exceeded.

Lo CAP A/D has gone below 2000 raw counts.

– – – – –

Middle dashes indicate that the display is attempting to show a negative number greater

than five digits.