Page 78 MSI-9750A RF Remote Indicator • User Guide

9750A

HANDHELD RF REMOTE INDICATOR for

C

ELL

S

CALE

®

MSI Boot Loader Version 3.K3

(1) Update Application
(2) Change Baud Rate
(ESC) Start Application

12) After all sectors are programmed, press the

POWER

key on the 9750A. The 9750A will turn off. Remove

the serial cable from the 9750A Comm Port.

13) Turn the 9750A back on. The 9750A display will read “CLEARING” while it initializes all setup regis-

ters.

14) Restore display and other system parameters as needed. RF Modem network parameters for the last RF

Network used are not usually affected by the bootloading process. However, multiple RF Network settings

will have to be reprogrammed.

9750A SETUP DUPLICATING

When multiple 9750A’s are used for similar purposes, you can program all screens, function keys, strings, etc., on

just one unit. Then using the following procedure, copy the confi guration to the other 9750A’s. Since confi guring

9750A’s can be quite complex, this feature can save a tremendous amount of time.
First you must obtain a 9-pin ‘D’ null modem cable. A null modem cable will swap transmit and receive lines.

However, it is important that pin 9, which is connected to the battery, is not cross-connected. Remove any con-

nection on pin 9. Source Code version must be the same in both units. If not, update the source code fi rst.

1) Disable any Auto Printing functions on both 9750A’s.
2) Confi rm both Comm Ports are set to the same baud rate. Defaults are 9600 Baud, 8 data bits, 1 stop bit,

and no parity. The fi le is small enough that 9600 baud is plenty fast. Turn both units OFF.

3) Plug the Null Modem cable into both units. MAKE SURE PIN 9 IS NOT CONNECTED.
4) Turn on both 9750A’s. Wait for 10 seconds.
5) On the source unit (the one already confi gured), press

SETUP

,

SEND/PRINT

,

[4]

,

[4]

,

[4]

.

6) On the receiving unit, press

SETUP

,

SEND/PRINT

,

[4]

,

[4]

,

[4]

,

EXIT

(Display key).

7) If the receiving unit is not displaying “MSI9750A-5” press POWER followed quickly by ESC.
8) Press

ID

on the receiving unit.

9) Press

[6]

on the source unit. Receiving unit should say “LOADING”.

10) The source unit will display percent done.

When fi nished the receiving unit should reset. It may also ignore its fi rst key press.

Modem settings are not transferred to prevent netwo

Modem settings are not transf

Modem settings are not transf

rk conflicts. Make sure all units network settings are configured properly.

rk conflicts. Make sure all units netw

rk conflicts. Make sure all units netw