6

Summit 3000 Installation Manual

2.6

Pit Installation

The Summit 3000 can be installed in a shallow pit using the optional Summit 3000 pit frame. Optional

height-adjustment holes are available. The following site considerations and pit frame drawings are meant only

as a brief overview of the principles involved with mounting the scale in a floor-level pit. The pit must be

installed in a suitable poured-concrete foundation according to standard construction practices.

Figure 2-5. Pit Frame Diagram

Site Considerations

Debris, floor sweepings, or material spills may accumulate in the pit and interfere with scale operation. You

should periodically clean the pit. All sizes of the Summit 3000 have a threaded hole centrally located for eyebolts

so the scale can be easily lifted from the pit.
Weight overloads, even momentary ones such as driving a loaded forklift over a scale corner, will damage load

cells. Plan the pit location out of main traffic areas to prevent such accidental damaging overloads.
A 1:24 slope is recommended for the pit, with full grouting under the corner pads as shown in Figure 2-6.

Figure 2-6. Section Showing Installed Summit 3000 and Pit Frame

2.7

Access Ramps

Access ramps for Summit 3000 floor scales are designed to bolt to the floor, with built-in mounting plates that

attach to the scale feet. When used with access ramps, side movement of the scale is automatically eliminated,

and no other mounting plates are necessary.

COVER PLATE

1:24 SLOPE

LOAD CELL

CUP

FOOT ASSEMBLY

GROUT