Installation

5

Figure 2-3. Junction Box Wiring Diagram

7. The cable must be routed to the indicator in a manner that will protect the cable from damage. This

method of cable protection in non-washdown applications are shown in Figure 2-4. When planning cable

routing, leave a loose coil of excess cable under the scale to facilitate future lifting of the scale for

servicing or cleaning.

Figure 2-4. Load Cell Protection

8. When the interface cable is protected and in its final position, complete connections to the indicator. See

the indicator installation manual for wiring information.

9. If necessary, trim corners as described in Section 3.2.
10. Check all strain relief fittings for tightness.
11. Put the cover back onto the junction box assembly and place the junction box back into the floor scale

cutout.

12. Secure the Summit 3000 floor plate.

Cable Color Code

Junction Box

Red

+ Excitation

Black

- Excitation

Green

+ Signal

White

- Signal

Bare

Shield

Yellow

+ Sense

Blue

- Sense

Table 2-1. Junction Box Connections

I n d i c a t o r
Te r m i n a l
Location

FLOOR

LOAD CELL CABLE

STEEL COVER