Instruction Manual - TORBAL AGZN Series

1

3

5. The scale will display “CAL”, indicating the calibration adjustment has been entered.

6. Before calibration adjustment can begin, the scale will perform automatic tare. Dashed lines “- - - - - - -” will

be displayed during auto tare.

7. Once the scale has been tared, the command “LOAD” will be displayed. At any time during the “LOAD”

command, place the single calibration weight in the middle of the pan.

8. Command “CAL” will be displayed, indicating that calibration is in progress.

5 Calibration can last from 15 to 30 seconds. When it is finished the scale will return to the weighing mode

2

e the scale returns to the weighing mode, remove the weight from the pan.