Instruction Manual - TORBAL AGZN Series

Chapter 15.

PRINTING AND RS232 COMMUNICATION PORT CONFIGURATION

After a weighing transaction is completed, a result data receipt can be printed. To initiate printing, wait for the
weighing result to stabilize and press the P key.

15.1.1.Torbal Printer Configuration (RXP-4)

The printing function is used to print data via the RS232 port. Data may be printed via the printer or sent to a PC via
the Torbal Communication Software.

To connect the AGZN Scale to a Torbal (RXP-4) printer, use the following settings:

Scale settings:
Baud rate: 9600
Parity: none
Data bits: 8

Printer settings (follow the steps detailed in the RXP-4 instruction manual):

Baud rate: 9600
Data: 8 bits
Parity: none/ignore
Column printing: 40

RS232 Communication Port Configuration