Message codes – Leica Geosystems Leica DISTO Pro User Manual

Page 59

Advertising
background image

61

DISTO™ pro

4

/ pro

4

a-1.0.0en

English

en

Repeat these tests at frequent and

regular intervals, also before and

after important measuring jobs.

Affix an adhesive label on the DISTO

for accuracy tests of measuring

equipment and keep a detailed

record of the test procedure.

Your DISTO meets the specified

accuracy if standard deviation

remains smaller or equal to the

typical specified value.

A DISTO whose measuring accuracy

has been tested over a test distance

works within the specified tolerance

over the entire distance and

temperature range specified in the

manual.

Please note the technical data and

the description of measuring

accuracy in the manual

Message Codes

Press briefly to acknowledge

messge and to return to basic

settings.

When messages with this

symbol, appear even after

DISTO has been used

several times, then please

call the service station and

inform them of the message

number displayed.

Error no.

Cause

Remedy

E702 - E706

Calculation error

Repeat entry

E252

Temperature above 50°C

Cool down instrument

E253

Temperature below 0°C

Warm up instrument

E255

Receiver signal too weak,

Measurement time too long, distance < 250 mm

Use target plate

Measurement time > 10 sec.

E256

Receiver signal too powerful

Use target plate (correct side)

E257

Wrong measurement, ambient brightness too high

Use target plate

E504

Not in data memory

No data available

E505

Data memory full

Erase memory location

E...

All other messages

Call service "system"