4 S

ETS OF

C

OMMANDS IN

O

FFLINE

M

ODE

(S

TANDARD

-M

ODE

)

4.1 ON-command (a)
Function:

Power on or reset the DISTO

TM

pro

4

/ DISTO

TM

pro

4

a. Same function as com-

mand “c“

Input:

a<cr>

Output:

?<cr><lf>

or @Exxx<cr><lf>

4.2 Online-command (EXT or A)
Function:

Switches on the extended set of commands (online mode) and switches to
online mode (all applications are off in this mode).

Input:

EXT<cr> or A<cr>

Output:

?<cr><lf>

or @Exxx<cr><lf>

4.3 OFF- command (b)
Function:

Switches off DISTO

TM

pro

4

/ DISTO

TM

pro

4

a .

Input:

b<cr>

Output:

?<cr><lf>

or @Exxx<cr><lf>

Remarks :

Short waiting period of about 500ms after switching off before they can be
switched on again.

4.4 STOP/CLEAR- command (c)
Function:

Stops current distance or signal measurement.

Input:

c<cr>

Output:

?<cr><lf>

or @Exxx<cr><lf>

4.5 Distance measurement (g)
Function:

Triggers single distance measurement.

Input:

g<cr>

Output:

WI31 WI51<cr><lf>

or @Exxx<cr><lf>

4.6 Tracking (h)
Function:

Triggers continuous distance measurement. Measurement is continued until
next command or until an error occurs.

Input:

h<cr>

Output:

WI31 WI51<cr><lf>

or @Exxx<cr><lf>

Remarks:

Interface must be set to 9600 baud or higher to fully exploit the measuring
rate.

4.7 Signal measurements (k)
Function:

Triggers continuous distance measurement. Measurement is continued until
next command or until an error occurs.

Input:

k<cr>

Output:

WI53<cr><lf>

or @Exxx<cr><lf>

Remarks:

Interface must be set to 9600 baud or higher to fully exploit the measuring
rate.

4.8 Laser (o, p)
Function:

Switch laser on/off.

Input:

o<cr>

Switch laser on.

Advertising