Instruction Manual MM850481 ed.01

6 CONNECTIONS

6.1 Electro-pneumatic

connections

To prepare the instrument for operation connect the sampling probe to the instrument:

• Plug the pneumatic connector of the probe to the GAS INLET connector. If you are using the gas sampling probe

with Draft option, connect also its connector to the instrument Pressure/Draft inlet.

• Plug the probe temperature sensor plug to the pertinent connector.

• For more accurate readings and efficiency calculations, plug the remote air temperature Pt100 sensor to the 3 pin

connector on the left side of the instrument.





pneumatic ports

Right side

P1 – Draft/Pressure

input

INLET – gas input

P2 – differential
pressure input

Tgas input

IR serial interface

RS232 – External
probes input

Battery charger

Pt100 Tair input

6.2

Gas probe positioning

The measurement site should be arranged at a distance of 2xD behind the exhaust gas connection pipe, where D =
diameter of the exhaust gas connection pipe.
To position the sampling probe in the exhaust gas pathway a hole of 11/16mm should be drilled and the retaining cone of
the sampling probe firmly screwed in it.
The retaining screw in the cone enables the probe to be easily moved in order to locate the core flow, normally
correspondent to the center of the section of the stack.
The flue gas and exhaust gas pathways, must be checked for gas-tightness before carrying out a measurement and, if
necessary, non gas-tight points should be sealed.
To locate the gas core flow, insert progressively the sampling probe and read the maximum value of temperature.