Ambient co probe (option), Co probe operations, 10 ambient co probe (option) – Eclipse Combustion EGA4 User Manual

Page 29: 1 co probe operations

Advertising
background image



Instruction Manual MM850481 ed.01

7.10

Ambient CO probe (option)

This optional external probe allows ambient CO monitoring for personnel safety. When connected to the analyzer, it
monitor continuously CO concentration and it signal when the Long-term Exposure (TWA) limits are reached. An
acoustic and visible signals are emitted by unit.
The following page show the current exposure limits for CO gas. This table is a guidance only, and figures must be
checked before being used. Sensors used in safety applications should be fully calibrated at least every six months.

U.K.
HSE

GERMANY

DFG

USA

OSHA

TWA (8h)

50 30 35

STEL (10’)

300 60

(30) 200

7.10.1

CO probe operations

To enable CO monitoring press the [MENU] key from measuring page. The following page will be displayed:

MENU

MEASURE

RECALL MEMORY

FLOW/RETURN TEMP

AMBIENT ANALYSIS

ESCAPE

Use the [S] and [T] keys to move the cursor symbol to “MEASURE” option.

Press the [ENTER] key to select the option. Press [ESCAPE] (F2) to return to the previous menu.

MEASURE

P DECAY LEAK TEST

GAS LEAK TEST

AMBIENT CO

ENVIRON. TEMP/HUM

ESCAPE

Use the [S] and [T] keys to move the cursor symbol to “AMBIENT CO” option.

Press the [ENTER] key to select the option. Press [ESCAPE] (F2) to return to the previous menu.


Connect the probe to the unit and wait for the following page:

AMBIENT CO

Amb. CO: 7 ppm

ESCAPE

ZERO


NOTE: W

HEN PROBE IS CONNECTED THE INSTRUMENT BEGIN A

40

SECONDS WAITING TIME

. I

F PROBE IS CONNECTED IN

A

MBIENT

CO

MENU

,

A WAITING MESSAGE WILL BE DISPLAYED

.


Before insert the probe in working area, agitate it for same seconds in fresh and uncontaminated air. Press the [ZERO]
key to reset the measure and start CO monitoring.
From this moment the unit will compare the monitored value with the CO limit (default OSHA limit: 35ppm). It is possible
to change the CO limit using the GasConfig software for PC.
Press [ESCAPE] (F2) to return to the previous menu.

NOTE: I

F THE PROCEDURE ABOVE IS EXECUTED AND THE

CO

PROBE REMAIN CONNECTED

,

THE INSTRUMENT WILL MONITOR AMBIENT

CO

ALSO DURING OTHER OPERATIVE MODE

.


If alarm condition is marked, press the [ENTER] key to silence it.

29