Eclipse Combustion EGA4 User Manual

Page 28

Advertising
background image



Instruction Manual MM850481 ed.01

Use the [S] and [T] keys to move the cursor symbol to the value to be changed. Press the [ENTER] key to store the
new selection.


Select “TEST RESULT” option for results viewing and press the [ENTER] key.

Now you can connect the instrument Pressure inlet (INP) connector to the gas circuit.

CAUTION

BE SURE

FOR EMPTY AND CLEAN CIRCUIT

:

OVERPRESSURE

,

DUST

,

LIQUID AND CONDENSING MAY DAMAGE THE

INSTRUMENT INTERNAL CIRCUITS

.



With circuit depressurized, press the [TEST] key. The unit will display the following page:

SELECT TEST

0.2 mmH2O

Result: A
START TEST

ESCAPE

ZERO


Press the [ZERO] key to reset the reading.

The 4000 series can memory store up to 2 different test data (A and B) for “Before” and “After” check.

Use the [S] and [T] keys to move the cursor symbol to “Result” option. Press the [ENTER] key to select the option.

Use the [S] and [T] keys to move the cursor symbol to the value to be changed. Press the [ENTER] key to store the
new selection.


Select “START TEST” option and press the [ENTER] key to run the decay leak test.

P1 : 100.0 mmH2O

P2 : 99.5 mmH2O
P3 : _____
P4 : _____

~P : _____

Result A: _____ 300

ESCAPE


The unit will store 4 different measurements at programmed time (standard values are: 10, 60, 120 and 120 second).
See paragraph 10.2.4 to change these values.
The instrument will display the

∆P (P1-P4) value and it compute, using the circuit specifications (Length, Diameter and

Type) the test result (GOOD or FAIL).
Test result will be calculated using the following acceptance limits table:

Under track circuit

Visible circuit

Diameter

[in]

½”

¾”

1”

1” ¼

1” ½

½”

¾”

1”

1” ¼

1” ½

Length

< 10m

3.5 3 2.75 2 1.75 35 30 27.5 20 17.5

< 20m

2.85 2.35 2.2 1.4 1.1 28.5 23.5 22 14 11

< 30m

2.7 2.2 1.5 1.1 0.85 27 22 15 11 8.5

< 40m

2.35 1.7 1.2 0.9 0.4 23.5 17 12 9 4

< 50m

2.1 1.4 1.1 0.3 0.3 21 14 11 3 3

> 50m

1.85 1.4 0.9 0.3 0.3 18.5 14 9

3 3

Note: These values are expressed in mmH

2

O. The plant circuit is failed when (P1-P4) exceed the values after 5

minutes.

28