Accuracies and ranges, 2 accuracies and ranges – Eclipse Combustion EGA4 User Manual

Page 7

Advertising
background image



Instruction Manual MM850481 ed.01

1.2

Accuracies and ranges

Parameter

Sensor type

Range

Resol.

Max

response

Accuracy

O

2

Electrochemical

from 0 to 25.0%

0.1%

20 sec.

±0.1% vol.

CO
H

2

compensated up

to 1000ppm

Electrochemical

from 0 to 8000 ppm

1 ppm

50 sec.

±10 ppm <300 ppm
±4% rdg up to 2000 ppm
±10% rdg >2000 ppm


CO

Electrochemical

from 0 to 20000 ppm

1 ppm

40 sec.

±10 ppm <300 ppm
±4% rdg up to 2000 ppm
±10% rdg >2000 ppm

NO

Electrochemical

from 0 to 4000 ppm

1 ppm

40 sec.

±5 ppm <125 ppm
±4% rdg >125 ppm

NO

2

Electrochemical

from 0 to 1000 ppm

1 ppm

50 sec.

±5 ppm <125 ppm
±4% rdg >125 ppm

NO

x

Calculated

from 0 to 5000 ppm

1 ppm

SO

2

Electrochemical

from 0 to 4000 ppm

1 ppm

40 sec.

±5 ppm <125 ppm
±4% rdg >125 ppm

CO

2

Calculated

from 0 to 100%

0.1 %

Tair

Pt100

from -10 to 100 °C

0.1 °C

±(0.2% rdg+0.15°C)

Tgas

Tc K

from -10 to 1000 °C

0.1 °C

±(0.3% rdg+0.3°C)

T

Calculated

from 0 to 1000 °C

0.1 °C

Tflow

Tc K

from -10 to 100 °C

0.1 °C

±(0.3%rdg+0.3°C)

Treturn

Tc K

from -10 to 100 °C

0.1 °C

±(0.3% rdg+0.3°C)

Pressure/Draft &
P

Piezo

-10hPa to +100 hPa

0.01 hPa

±1% rdg

Excess air

Calculated

from 1.00 to infinite

0.01

Efficiency

Calculated

from 0 to 100%

0.1 %

Gas speed

Pitot tube

From 0 to 100.0 m/s

0.1 m/s

Smoke index

External pump

from 0 to 9

Technical units and ranges can be converted directly from ppm to mg/Nm

3

, mg/kwh and from hPa to mmH

2

O, mbar or inH

2

O.

The relative accuracy shown are expressed as absolute or % of rdg errors at -5°C to +40°C ambient temperature.

The maximum response time shown is referred to 90% signal changes.

The pressure relative accuracy shown is valid only after the auto-zero procedure.



Specifications may change without notice.