Flue gas analysis, Ecoline 4000, Greenline 4000 – Eclipse Combustion EGA4 User Manual

Page 21: Unigas 3000, 2 flue gas analysis

Advertising
background image



Instruction Manual MM850481 ed.01

7.2

Flue gas analysis

Connect the gas probe to the analyzer as shown in chapter 7. DO NOT INSERT THE PROBE IN THE CHIMNEY.

Switch the analyzer on pressing the [ON] key. The display will show the following page:

EcoLine 4000


Num. Serie:

1257

Versione: 1.000


GreenLine 4000


Num. Serie:

1257

Versione: 1.000


UniGas 3000+


Num. Serie:

1257

Versione: 1.000


This page changes automatically after same seconds or if you press a key as follows:

! Don t insert !
! probe in chimney !

Op.: Antonio
Bat: 5.2V (23.5°C)
Autozero: 180 sec

MENU

MEASURE


This page displays the following information: operator name, battery charge level, instrument internal temperature
and the seconds to end the autozero procedure.


N

OTE

: T

HE ANALYZER MUST DRAW CLEAN AIR FROM THE AMBIENT DURING THE AUTOZERO PROCEDURE

. O

THERWISE THE

MEASUREMENTS CANNOT BE ACCURATE

.

It is possible to enter in “MENU” page (pressing [F1]) or in “MEASURE” page (pressing [F2]). See the appropriate
paragraphs for operations.
When autozero procedure finished, an intermittent sound advise the operator. Press the [ENTER] key to mute the
unit. The following page will be displayed:

! Now insert !
! probe in chimney !

--- FUEL --

Natural Gas

TAG SELECTION

ANALYS.


This page allows the operator to change the fuel used and to select the Tag between the available.

Fuel selection. Move the cursor symbol “X” on the desired row using the [S] and [T] keys. Press [ENTER] to

change the value. See instruction on paragraph 8.1.6.

Tag selection. Move the cursor symbol “X” on the desired row using the [S] and [T] keys. Press [ENTER] to

change the value. A new page will be shown. The Tag name, the description and the progressive number will
be displayed. See instruction on paragraph 8.1.6. When in measuring mode, you can select another Tag
pressing the [MEMORY] key.

Insert the probe in chimney. Stop it using the retaining cone and adjust the immersion depth for a gas sampling in
the middle of the exhaust gas connection pipe.

Press the [ANALYSIS] (F1) key to begin the measurements. The following page will be shown:

O2 20.9 %

CO2 0.0 %
CO 1 ppm
NO 1 ppm

Tf Tc Error
Eff --------

Ref O2

GRAPHIC

21