Pet ether v <enter, Close vessel water on <start, Select temp 90 (^v)... <enter – ANKOM XT15 Series User Manual

Page 18: Select time 60 (^v)... <enter, Operator’s manual

Advertising
background image

Operator’s Manual


pg. 18

Rev F 12/08/14

3.6 Slide the Extraction Vessel into the instrument ensuring the

Extraction Vessel O-ring stays in place.

3.7 Pull the Main Handle downward, closing and locking the Extraction

Vessel in place. You will hear a click noise when the Safety Valve
Pin locks in place.

Before running the instrument, verify that the Extraction Vessel is locked in
place by attempting to rotate the Main Handle upward. If it is not locked,
repeat step #8 above.

3.8 Press ENTER on the Keypad.

The Display should now show the screen to the right. (The
time shown will be the last time entered.)

4. Set the Extraction Time.

4.1 Using the arrow keys on the Keypad, select the desired Extraction time. (The recommended time is

60 minutes.)

4.2 Press

ENTER on the Keypad to set the Time.

The Display should now show the screen to the right. (The
temperature shown will be the last temperature entered.)

5. Set the Extraction Temperature.

5.1 Using the arrow keys on the Keypad, select the desired Extraction Temperature. (The

recommended maximum temperature is 90°C.)

5.2 Press

ENTER on the Keypad to set the Temperature.

The Display should now show the screen to the right.

6. Verify that the water supply is on.

7. Run the instrument.

Press

START on the Keypad.

The instrument will run through its automatic operation, showing
its status on the Display. When the instrument has completed its
automatic operation, the Display will show the screen to the right.

At this time, the only light shining on the Display should be the green
power light.

PET Ether
v <ENTER>

Close Vessel
Water On <START>

Select temp 90
(^v)... <ENTER>

Select time 60
(^v)... <ENTER>