Diagnostics, Heater (q0) off enter to toggle, Heater (q0) v <enter – ANKOM XT15 Series User Manual

Page 23: Outputs <enter> exit <stop> v, Diagnostics ^ <enter, Pet ether v <enter, Operator’s manual

Advertising
background image

Operator’s Manual


Rev F 12/08/14

pg. 23

Diagnostics

To help solve problems that may occur while running the ANKOM

XT15

Extractor, the following computer controlled

Diagnostics are available and accessible through the Display and Keypad.

1.

Toggle Outputs (Q0 Heater Relay and Q1 - Q7 valves)

2.

Check Temperature and Level Sensors

3.

Check the state of the ETS (Emergency Temperature Shut-off switch) relay

Access Diagnostics Screen

When the instrument is powered up or when the instrument finishes a run, the
screen to the right is displayed.

The factory default solvent shown on the Display is "Pet Ether." However, the
instrument computer will store in memory the last solvent selection made and use
it as the default solvent shown on the Display screen every time the instrument is
ready for a new run.

To access the Diagnostics Screen, press the Down Arrow button (

v) on the

Keypad four times. The screen to the right will be displayed. From the
Diagnostics Screen you can access all of the available diagnostics.

Toggle Outputs

To toggle outputs, follow the procedure below.

1.

With the Diagnostics screen displayed, press the

ENTER button on

the Keypad. The screen to the right will be displayed.

If you press the

STOP button on the Keypad while this screen is displayed, you

will go back to the Diagnostics screen. If you press the Down Arrow button (

v) on

the Keypad while this screen is displayed you will go to the next group of
diagnostics (Analog).

2.

With the Outputs screen displayed, press the

ENTER button on the

Keypad. The screen to the right will be displayed.

3.

Press the Down Arrow button (

v) on the Keypad to select the

heater relay (Q0) or valve (Q1 - Q7) you want to toggle. With
your selection made (in this example we selected the heater relay
Q0), press the

ENTER button on the Keypad. The screen to the

right will be displayed.

Heater (Q0) OFF
ENTER to toggle

Heater (Q0)
v <ENTER>

Outputs <ENTER>
Exit <STOP> v

Diagnostics
^ <ENTER>

PET Ether
v <ENTER>

Advertising