Check the state of the ETS (Emergency Temperature Shut-off switch) relay

To check the state of the ETS relay, follow the procedure below.

1.

Access the Diagnostics screen (see the "Access Diagnostics
Screen" section for details).

2.

With the Diagnostics screen displayed, press the

ENTER button on

the Keypad. The screen to the right will be displayed.

3.

Press the Down Arrow button (

v) on the Keypad. The screen to the

right will be displayed.

4.

Press the Down Arrow button (

v) on the Keypad. The screen to

the right will be displayed.

5.

Press the

ENTER button on the Keypad. The screen to the right

will be displayed showing the state of the ETS relay. If the ETS
relay is "OFF", contact ANKOM Technology.

6.

Press the

STOP button on the Keypad to go back to the Inputs

screen (shown to the right).

7.

Press the

STOP button on the Keypad to go back to the main

Diagnostics screen (shown to the right).

8.

Press the Up Arrow button (^) on the Keypad to exit Diagnostics.
Press the Up Arrow button (^) on the Keypad up to three more
times to return to your solvent screen (example shown to the
right).