Cleaning the water filter

To clean the water filter (located on the water supply line), follow the procedure detailed below.

1. Unscrew the filter from the instrument.

2. Unscrew the filter clamp from the water supply line.

3. Use compressed air to blow backwards through the

filter (blow opposite the arrow on the filter). If the
filter is clogged, replace it with ANKOM part # X97.

4. Reconnect the filter and fittings.

5. Turn on the water and check for leaks.

Replacing the Fuses

To replace the fuses in the ANKOM

XT15

Extraction System, follow the procedure detailed below.

1. Turn off the instrument power and unplug the Power Cord from the outlet.

2. Remove the two screws on the Top Cover.

3. Swing the Top Cover down exposing the electrical compartment.

4. Locate the Fuse Holders on the terminal strip. There

are two fuses within the instrument; one for each pole
(100 – 120V 10A or 220 – 240V 5A).

5. Press the tabs to lift the Fuse Holders up.

6. Replace the fuses and close the Fuse Holders completely.

7. Close the Top Cover and secure it with the two screws that you previously removed.

Replacing the Charcoal Filter

The Charcoal Filter should be replaced every 3 months, or after 100 extraction operations using the instrument. To
replace the filter, remove the old one from the vent line and install the new one.

Fuse
Holders

Tab to open
Fuse Holder

Fuse

Filter