Brookfield CT3 Texture Analyzer User Manual

Page 12

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 12

Manual No. M08-372-C0113

II.

QUICK START

1.

Unpack the instrument according to Section I.5.

2.

Install base table in accordance with the instruction sheet that is enclosed with the base

table. Place the sample on the base table. Adjust the table height so that the surface of the
sample is within 5 mm of the probe.

3.

Attach the selected probe. See Section IV.2 for more information.

4.

Set the test mode to Normal. Please review section III.6 for detailed explanation of operation
of Select/Scroll knob.

5.

Set the trigger value as recommended below.

Load Cell

Recommended Minimum

Trigger Value

100g

0.5g

1000g

2g

1500g

3g

4500g

4.5g

10kg

10g

25kg

25g

50kg

50g

Table II.1

6.

Set the test speed and distance. See Section IV.3 for more information.

7.

Press the START button. The weight of the probe will autozero and then the test will
start.

8.

Record all test results.

9.

Remove sample and clean the probe.