Trigger Point:

Range

Resolution

Model

Grams

Grams

CT3-100

0.1-10

0.01

CT3-1000

0.2-100 0.10

CT3-1500

0.2-150 0.2

CT3-4500

0.5-500 0.5

CT3-10KG

1-1000

1.0

CT3-25KG

2-2500

2.0

CT3-50KG

5-5000

5.0

Recommended trigger value settings are given in Table II.1

NOTE: Setting a trigger value of zero disables trigger function so test will immediately start

from current position.

Speed: 0.01 to 0.1 mm/s in increments of 0.01 mm/s

0.1 to 10 mm/s in increments of 0.1 mm/s

Accuracy: ±0.1% of set speed

Position:

Range: 0-101.6 mm

Resolution: 0.1mm

Accuracy: 0.1mm

Increments for changing test speed using the Select/Scroll knob can be switched between
0.1mm/s and 1.0mm/s by depressing Reset/Stop button.

Temperature:

-20°C to 120°C (Requires optional probe, DVP-94Y)

Output:

RS232 Compatible Serial Port, USB Port

Environmental

0°C to 40°C temperature range (41°F to 104°F)

Conditions:

20% - 80% relative humidity, non-condensing atmosphere

Use:

Intended for indoor use only

Altitude: up to 2000m

Dimensions:

10.5” x 10.5” x 24”

Weight:

37 lbs. (16.8 kg)

I.5 Installation

1) Prepare a clean, level surface.

NOTE: This instrument is a sensitive force measuring device. It should be installed on a

clean, solid, level bench surface which is free from external vibrations.

2) Unpack and remove the CT3 Texture Analyzer from the shipping container.

The CT3 Texture Analyzer weighs 16.8 kg (37 pounds). Use caution when lifting the
unit out of the packaging.

3) Place the CT3 Texture Analyzer on a sturdy, level surface. Adjust the four feet to ensure

that the instrument is stable.