Preparing the minipv series for testing – Cannon Instrument miniPV-X User Manual

Page 21

Advertising
background image

15

CANNON

®

miniPV-Series Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.a— January, 2011;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Compare the values in the

Tray Settings: Advanced

window with your

archived values and make any necessary changes; then click

OK

.

Preparing the miniPV Series for testing

Inserting the miniPV Series temperature probes/reference thermometer

Inserting the probes

The temperature probe and thermal fuse probe are semi-permanently
installed. Initial installation or replacement requires removal of the top
cover of the bath unit. The probes are inserted through the narrow
aperture on the top bath flange and held in place with a sliding metal plate
(friction fitting) secured with a thumbscrew.

miniPV Series top bath flange

Inserting the thermometer

The ASTM reference thermometer/adapter assembly can be inserted
through the grommet in the bath top cover and through the larger circular
aperture on the top bath flange (see photo image above). The assembly
includes a stop collar with set screw that can be adjusted to vary the
thermometer height.

CAUTION

Do NOT insert a reference thermometer while the miniPV Series bath
impeller is operating, as damage to the thermometer and/or viscometer
tube may result. Make certain to insert the appropriate reference thermom-
eter for the current bath temperature! Use appropriate safety procedures
when handling the thermometer, as it contains mercury.

Advertising