Cannon Instrument miniPV-X User Manual

Page 39

Advertising
background image

33

CANNON

®

miniPV-Series Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.a— January, 2011;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CAUTION

Make certain to enter the correct check standard viscosity (in centistokes)
BEFORE running the standards. Incorrect data entry may skew future
calibration results.

5. Click

Run

, then select the appropriate sample trays to test the

standards.

6. When testing is completed, use the VISCPRO

®

software to log on as

a Manager (

Main/Log In

) in order to access the calibration menu.

7. Select the desired miniPV Series instrument from the

Configure

menu; then select

Calibration

.

The

Calibration

window will open.

8. Select the date range for acceptable calibration data by using the

No

older than [x]

days spin controls

to set the date search param-

eters (if “

1

” is selected, only data from the standards run in the last

24 hours will be listed). Any check standards from the time frame
you specified which are appropriate for calibration calculations will
appear in the window list box.

The miniPV Series Calibration window

9. Use the spin controls

to select the desired

Bulb

for calibration.

Advertising