To adjust the vacuum/pressure settings for the miniPV Series, click

Service

from the VISCPRO menu options and select the desired instru-

ment group and instrument. Then click

Adjust Vacuum/Pressure

to

open the

Adjust Vacuum/Pressure

window.

Select the desired adjustment (vacuum or pressure) and then follow the
screen prompts to complete the adjustment procedure, using the appropri-
ate controls on the instrument wing panel. The prompts will display
recommended vacuum/pressure settings and will also indicate how
adjustments may be made to alter the settings.

To calibrate for temperature, see Chapter 3 for details. Temperature
calibration requires a reference thermometer with an accuracy meeting
the precision specification of ASTM D445.

To zero temperature offsets, click the

Service/Clear Temperature

Offsets

menu option for the desired instrument. The

Zero Temperature

Offsets

window will open. Click the check box to select each desired

action. Then click

OK

to clear the selected settings.

If the

Clear Current Offset

option is checked, VISCPRO will clear the

calibration point for the current bath temperature.