Checking wash functions – Cannon Instrument miniPV-X User Manual

Page 23

Advertising
background image

17

CANNON

®

miniPV-Series Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.a— January, 2011;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

3. Select the desired unit of temperature measurement (°C or °F).

4. Type the desired temperature in the

Bath Temp:

field. Acceptable

values are any numbers between 20°C and 100°C. If necessary, you
may use the decimal point key to input temperature to the nearest
0.01°C. Press

OK

to save the temperature setting.

NOTES

The miniPV Series will not allow sample testing until the instrument has
equilibrated temperature at the test temperature.

To cancel temperature selection before completing the procedure, click
Cancel.

After the target temperature has been set, the miniPV Series bath
temperature will be adjusted to the target temperature and the bath will
equilibrate at the test temperature. The VISCPRO II software always
displays the current and target bath temperature in the

Sample Input

wi

ndow.

Checking wash functions

To ensure function of the solvent wash system, follow the procedure
below:

1. Place an empty vial in the sample tray position 1.

2. Select

Wash viscometer tube

to open the

Wash viscometer tube

window.

Advertising