36

CANNON

®

miniPV

®

-HX Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.0— January , 2011;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

t

=

flow time (seconds)

NOTE

When calibrating the miniPV Series, the operator should analyze the
viscosity standard using double determinations to obtain two flow times for
each standard. The average of the two flow times is used as input data for
the calibration program.