completed. Sensor training data remains valid until the sensors are
retrained at the same temperature. Retraining is necessary only when the
viscometer tube is changed, or if sensor performance changes over time.

Train Tube Sensors

To train tube sensors, click

Service

menu and select the

Train Tube

Sensors

menu for the desired instrument. The

Train Tube Sensors

window will open.

Prewash option:

Click the

Prewash

option if the tube is not clean and

dry. VISCPRO will initiate a cleaning cycle using current wash configu-
ration settings prior to initiating the sensor training procedure.

Trip Points Only

option: Click the

Trip Points Only

option to adjust

the sensor trip points without setting the voltages. This is useful if it is
neccessary to adjust trip points on sensors for which voltages have
already been set.

After selecting the desired training options click

START

and follow the

software prompts to complete the sensor training.

Standard tube calibration

During standard calibration, the software uses check standard data
obtained from miniPV Series tests to perform the mathematical calcula-
tions necessary to derive new constants for each timing bulb of the
viscometer tube. Then the VISCPRO software displays the new con-
stants and computes a viscosity (in cSt) for each sample using the actual
flow time of the check standards. The resulting viscosity calculation is
compared with the known viscosity of the standard; absolute and com-
parative data is displayed on the computer screen.