Hardy Enviro Bench Scales User Manual

Page 19

Advertising
background image

Section 6: Serial I/O

12/09

14

0596-0315-01

Rev

B

X.xxupd: Display microvolts per division: This is a reference to display the current microvolts per division.
The display will indicate x.xxupd where x.xx is a numeric value of the microvolts per division. Press any
front panel pushbutton other than b/g / net to continue.

Acc: Accumulate feature enabled: Enables or disables the accumulate feature. The display will indicate acc
followed by the current setting of y for enable and n for disable. Press over/under/next, units or tare to
change the selection. Press Program/Enter to enter the selection.

A thld: Autoprint and accumulate threshold: This is the number of divisions the indicator must meet or
exceed for an autoprint or an auto-accumulation to occur. The display will indicate a thld, followed by the
current setting in divisions with the left digit flashing. Press units to increase the digit, press the 1 key to
decrease the flashing digit. Press Over/Under/Next to move the flashing digit to the right or press the tare to
move it to the left. Press Program/Enter when the display shows the correct threshold setting. The
acceptable range is 1 to 1000 and default value is 10 divisions.

Aout: Autoprint and autoaccumulate output: This setting will determine whether the indicator will autoprint
during an automatic accumulation. This setting is only active if port 1 is set to auto and accumulation is
enabled. The display will indicate aout followed by a y for enable or an n for disable. Press over/under/next,
units or tare to change the selection. Press Program/Enter to enter the selection.

Beep: Audible tone when key is pressed: This will enable or disable the audible tone to acknowledge pressing
a front panel key. The display will indicate beep followed by a Y for enable or an n for disable. Press
over/under/next, units or tare to change the selection. Press Program/Enter to enter the selection.

C-fact: Conversion factor for custom units: The primary units divided by this value will determine the custom
units if it is enabled. The display will indicate c-fact followed by a value with the left digit flashing. Press
units to increase the digit, press the 1 key to decrease the flashing digit. Press Over/Under/Next to move the
flashing digit to the right or press the tare to move it to the left. Press print to move the decimal point to the
right. Press Program/Enter when the display shows the correct conversion factor.

Note: An example of custom units might be to display the weight in tons. In this case the conversion factor

is 2000.0 if the primary unit is lb.

The default conversion factor is 8.3333.

The Enviro™ Series indicator will now return to the setup prompt. Navigate to the desired menu heading to
continue programming, or to done to finish.