3 multipoint calibration, Multipoint calibration – Hardy Enviro Bench Scales User Manual

Page 23

Advertising
background image

Section 6: Serial I/O

12/09

18

0596-0315-01

Rev

B

5.3

Multipoint Calibration

Press and hold the PROGRAM/ENTER key until “SEtUP is displayed.

Or press internal switch S1 to display “SEtUP”

Press the OVER/UNDER/NEXT key until “CAL” is displayed

Press the PROGRAM/ENTER key to display security level “SL”

Press OVER/UNDER/NEXT to select proper security level

Press PROGRAM/ENTER

Raw counts displayed

No

New Zero?

Yes

Press the ZERO key

Scale zero’d. Raw counts become zero reference

ZERO

Press

PROGRAM/ENTER

Displays “CALAt2”, “CALAt3”,

Displays “CALAt1”, then

or “CALAt4”, then calibration weight

a calibration weight value

Enter test weight value. Most Significant digit blinks.

Increase digit with UNITS key. Decrease digit with 1 key.

Move flashing digit left using TARE key, or right using OVER/UNDER/NEXT key.

ZERO key clears display, PRINT moves decimal to right.

Next page