Section 6: Serial I/O

12/09

17

0596-0315-01

Rev

B

Displays “LOAd1”, then raw counts

Apply test weights to scale. Press PROGRAM/ENTER key

Displays correct weight. Note: UNITS LED blinking

Press the 2 key or internal S1

Displays microvolts per division (x.xxuPd)*

Press any front panel key to return to “CAL” prompt

Press OVER/UNDER/NEXT to display “donE”

Press PROGRAM/ENTER to display “SAvE”

Press PROGRAM/ENTER to display “SAvEd”

Scale returns to weigh mode

*The

uPd”

(microvolts per division) must be one (1.0) or greater for proper operation.