Kistler-Morse KM Weigh II User Manual
Page 49
Chapter 6. Calibration
6-11
8. Press the F3 Key to access the Set
Menu. The display looks like this:
9. Use the keypad or the Up and Down Arrow
Keys to enter a new value for the Raw
Input. Press the Enter Key to save the
new value in memory. The display then
switches to a display that looks
like this:
10.
Use the keypad or the Up and Down
Arrow Keys to enter a new value for the
Corrected Output. Press the Enter Key to
save the new value in memory. The display
then switches to a display that looks like
this:
11.
Repeat Steps 9 and 10 as desired
until you have completed your setup of the
linearization table.
12.
Press the Esc Key to scroll up the
menu tree or press the Auto/Man Key to
return the display to vessel monitoring.
Indicated
Weight
Counts
○○○
○○○○○○○○
○○○○○○○○
○○○○○
Non-Linear
Input
1R
1C
2C
2R
4R
3R 4C
5R
5C
○○○○○○○○○○○○
○○○○○
Linearized
Output
Figure 6-4. Linearization Curve
Pt RAW INPUT
01> 1048576 Cnts
F1 F2 F3
Pt CORRECTED OUT
01> 1048576 Cnts
F1 F2 F3
Pt RAW INPUT
02> 1112576 Cnts
F1 F2 F3
Linearizing Table
Raw Input
Corrected Output
1R
1C
2R
2C
3R
3C
4R
4C
5R
5C
Requirements of table values:
1R<2R<3R<4R<5R
1C<2C<3C<4C<5C
1R=1C
5R=5C
3C