Chapter 7. Service

7-7

CAUTION

Manually activating setpoint relays
may cause damage if control equip-
ment is connected. Disconnect control
equipment before proceeding.

A typical setpoint test display looks like this:

• SP 02 is the current setpoint reference

number (referred to as SP2 when you
added it in the I/O Menu).

• SL 02 is the slot number of the setpoint

PCB.

• Ch2 is the channel number of the setpoint

on the setpoint PCB. Each setpoint PCB
has eight individual channels.

• On indicates the current On/Off status of

the setpoint. Pressing the F1, F2, or F3
Key toggles the setpoint between
On and Off.

Pressing the Up and Down Arrow Keys or the
Enter Key toggles the display between the
setpoints for the current vessel.

Pressing the Esc Key or Auto/Man Key
terminates the test. Once the test is
terminated, the setpoints return to
automatic control.

4/20

This menu allows you to calibrate the 0/4 and
20 mA outputs to the device that is receiving
the current output. It also allows you to set
the current output to specific values for test
purposes. The 4/20 Menu has two submenus:

Iadj

This menu is used to manually assign
counts-to-milliamps outputs for 0 mA, 4 mA,
and 20 mA outputs to calibrate to another
device. Follow this procedure:

1. Put the Weigh II in the Manual Mode, go to

the Main Menu, the Service Menu, the
4/20 Menu, and then the Iadj Menu. The
display shows:

2. Next (F1) scrolls the display forward

through all the current channels. Back (F2)
scrolls the display backward through all
the current channels. Scroll to the desired
current channel using the F1 or F2 Key.

3. Press the F3 Key to access Select. The

display shows:

Note

If you have set up 4-20 mA output, the
Weigh II ignores anything you enter for
0 mA. Similarly, if you have set up 0-
20 mA output, the Weigh II ignores
anything you enter for 4 mA.

4. Press a Function Key (F1, F2, or F3) to

assign counts to the associated current
output. If you select F1 (0 mA), the display
looks like this:

5. Use the keypad or the Up and Down Arrow

Keys to change the counts if desired.
Press the Enter Key to save the value in
memory. The display flashes a message
acknowledging your selection and returns
to:

6. If desired, repeat Steps 4 and 5 for the

other current output, and repeat Steps 2
through 5 for other current channels.

Test

This function allows manual activation of
current output channels outside of normal
control. The system issues a warning that the
automatic control of current outputs assigned
to the currently selected channel is trans-
ferred to manual control, and requests
verification that that is what you want to do.

CAUTION

Manually activating current output can
cause damage if control equipment is
connected. Disconnect control equip-
ment before proceeding.

SP 02 :SL 02:Ch2

TOGGLE STATE>ON

F1 F2 F3

CURRENT CHAN#: 1

Next Back Select

F1 F2 F3

SELECT CAL POINT

0mA 4mA 20mA

F1 F2 F3

0mA CAL POINT

> X cnt

F1 F2 F3

SELECT CAL POINT

0mA 4mA 20mA

F1 F2 F3