14. Connecting the Scale to the AD-8124 Label Printer

14-1 Confirming the SF/SG Program Version

Confirm the SF/SG program version as follows:

Step 1: While holding down the ZERO key ( ) of the numerical keypad , turn the power switch

on.

Step 2: When

“888888 ”

appears, the scale program version is earlier than 4.00. When “

rP XXX

appears, the scale program version is 4.00 or later.

Earlier than 4.00

4.00 or later

Note about Connection to the AD-8124

• To connect the SF/SG scale to the printer, use an RS-232C crossover cable (D-Sub 9-pin socket type) such

as an AX-KO1371-200 (2 m).

• To get optimum performance from the AD-8124, use the SF/SG program version

rP 4.00

or later.

Some functions can not be used with the earlier versions, such as the PLU number printing, printer internal
clock setting and scale PLU unit price data changing.

• The totaling function and multiplier function of the SF/SG scale can not be used.

• When an item is weighed and the data is sent to the printer, the next data transmission is not available until

the item is removed from the scale.

14-2 Setting the “F” Functions

Set the “F” functions depending on the scale program version as follows:

Program version 4.00 or later:

(1) Output mode

Label printer mode

F3=8

(2) Baud rate

9600 bps

F4=4

(3) Parity bit

8 bits (No parity)

F5=1

Program version earlier than 4.00:

(1) Output mode

Key trigger mode 1

F3=0

(2) Baud rate

9600 bps

F4=4

(3) Parity bit

8 bits (No parity)

F5=1

AD-8124 continued...

888888

888888

888888

WEIGHT

UNIT PRICE

PRICE

rP 4.00

WEIGHT

UNIT PRICE

PRICE