8.2 Automatic adjustment via Clock-CAL

(only models ABT 100-5M, ABT 120-5DM, ABT 220-5DM)

With the help of its internal adjusting weight and integrated clock the balance can be

set to carry out automatic adjustment at set times (up to three times daily). Clock-

CAL is a very convenient function, when calibration reports are desired to be made

for regular calibrations, or when wishing span calibrations during break times to avoid

interruption of measurement work.

The weight symbol blinks for about two minutes as notification of span calibration

before it begins. Automatic adjustment can be stopped by actuating the [ON/OFF]

key during this message.


Clock-CAL function ON/OFF:

Press the [CAL] key repeatedly until "SettinG" appears.

Press the

[TARE]

key. The display shows “CAL dEF”.

Actuate the [CAL] key repeatedly until the current "tCal:∗∗" setting

appears:

To change the setting, actuate the [TARE] key as soon as

"tCal:∗∗" appears.

You can use the [CAL] key to select between the following set-

tings:

“tCAL-on”

Function activated

“tCAL-oF”

Function deactivated

The current setting is indicated by the standstill display ( ).

Confirm your selection by pressing the

[TARE]

key

To exit the function, press the [ON/OFF] key.

Brief actuation of ON/OFF key:

Back to previous menu.

Long actuation of ON/OFF key:

Back to weighing mode.