Repeatedly press the [UNIT] key until the balance is in density

determination mode for solids ",d". Note that “g” also appears dur-

ing weight measurement in air.

Press the [TARE] key. Place the item to be measured on the

weighing pan.

When standstill control is complete, press the [CAL] key

Place the item to be measured on the immersed hanging weighing

pan. The display is showing the density of the measured item.

“dSP oL” may be displayed when nothing is on the weighing pan,

which is normal.

14.3.2 Determining density of liquids

Condition

: Function "d" (density of liquids) is activated. See Chapter 11.3.

Press the [CAL] key repeatedly until "SettinG" appears.

Press the

[TARE]

key.

Repeatedly actuate the

[CAL]

key until "Sv SEt" appears

(Example)

Press the [TARE] key. The currently set density for the body to be

immersed appears. In the upper part of the display panel, the

symbol and the # symbol appear in order to indicate numer-

ical input status. The leftmost digit blinks.

(Example)

Enter density for your body to be immersed. When the [UNIT] key

is pressed, the numerical of the blinking digit increases by 1 at a

time. You can determine the value of the flashing digit, or shift the

flashing digit by one position to the right, by pressing the [PRINT]

key. Confirm your setting by pressing the [TARE] key.

Repeatedly press the [ON/OFF] key until the balance is in weigh-

ing mode.

Advertising