KERN & SOHN ABT 220-5DM User Manual

Page 72

Advertising
background image

ABT-BA-e-1312

72

Attempting to measure volatile sub-

stances

Measure with a lid on it.

The weighed item is electrically

charged.

Measure in a metal contain-

er. Measure with a metal

object larger than the items.

Sample temperature and tempera-

ture inside weighing space differ.

Measure at the same tem-

perature. Leave the item in

the chamber before meas-

urement. Change to high-

stability mode.

Draft in weighing space.

Leave the glass doors of the

weighing chamber open 1 to

2 cm wide when not in use.

Effects of electronic noise or strong

electromagnetic waves

Move away from the noise

source.

Internal trouble with the balance

Contact a service repre-

sentative.

“oL” or “-oL” is dis-

played

The load on the weighing pan is too

large. The weighing pan is de-

tached.

Use within the weighing

capacity. Place the weighing

pan on correctly.

Automatic adjustment

carried out frequently.

Severe temperature variations in

the room or the instrument

Move to a location with less

temperature fluctuation.

Display is faulty.

Adjustment is not carried out.

Carry out correct adjust-

ment.

No taring to zero before weighing.

To reset the display before

weighing, actuate the

[TARE]

key.

The desired weighing

unit cannot be called

by [UNIT] key.

Unit was not activated beforehand. Set unit beforehand.

Cannot transmit or

receive data to or from

computer or device.

Communication settings are wrong. Make the proper communi-

cation settings.

Error message ap-

pears.

Refer to the error code ta-

ble.

During

adjust-

ment

Error message ap-

pears.

Refer to the error code ta-

ble.

During

menu

item se-

lection

The menu settings

cannot be changed.

The menu is locked

Remove the menu lock.