10 iso/glp log, 1 setting the log printout of your adjustment data – KERN & SOHN ABT 220-5DM User Manual

Page 31

Advertising
background image

ABT-BA-e-1312

31

10 ISO/GLP log

Quality assurance systems require printouts of weighing results as well as of correct

adjustment of the balance stating date and time and balance identification. The easi-

est way is to have a printer connected.

10.1 Setting the log printout of your adjustment data
This function enables automatic log issue after each adjustment. These logs may be

issued by using an optional printer. Fully automatic and periodical adjustments are

carried out and logs created in combination with the Clock-CAL function (see ch.

8.2).

First make adjustments as specified in chap. 15.4 under „KERN-YKB-01N“.

Then set the log printout as follows:

Press the [CAL] key repeatedly until "SettinG" appears.

Press the

[TARE]

key. The display shows “CAL dEF”.

Repeatedly actuate the [CAL] key until "GLP:**" appears (The ∗∗

symbolise the current setting).

To change settings, actuate the [TARE] key when "GLP:**" ap-

pears.

You can use the [CAL] key to select between the following set-

tings:

“GLP-on”

Function activated

“GLP-oF”

Function deactivated

The current setting is indicated by the standstill display ( ).

Confirm your selection by pressing the

[TARE]

key

Actuate the [ON/OFF] key repeatedly. Return to menu/weighing

mode.