5 add up mode – KERN & SOHN ABT 220-5DM User Manual

Page 60

Advertising
background image

ABT-BA-e-1312

60

14.5 Add up mode
An optional number of single weighing processes are automatically added up to a

total sum, e.g. all single weighing processes of a batch or if you have to carry out a

great number of measurements of small samples.

When the standstill control ( ) is complete the weighing value is automatically is-

sued to the optional printer. The displayed value is added into the total adding

memory. Afterwards automatic taring will take place. This is repeated newly every

subsequent time a new sample is placed on the weighing pan. When the last single

weighing process is finished, press the [ON/OFF] key to receive the total sum ("TO-

TAL=").

Important:

This mode can be applied to any unit.

The initial load of the balance must be kept within the ± 5-fold of the ze-

ro range.

Weight of sample ≥ 10 x zero range (zero range = 0.25 e)

If during the use of multiple range balances the smallest readability digit

was turned off by pressing the [1d/10d] key, evaluation is based on the

latest add-on value displayed before.


Activate function:

Repeatedly press the [CAL] key until "FUnC.SEL" appears.

Press [TARE] key

(Example)

Repeatedly press the [CAL] key until the current "Addon:**" set-

ting appears.

Press

[TARE]

key

You can use the [CAL] key to select between the following set-

tings:

Adon-on”

Function activated

“ Adon-oF”

Function deactivated

The current setting is indicated by the standstill display ( ).

Advertising